Check Room
Verify that the cartridges stored at room temperature are within the expiration date
Temperature
and that the cartridges have been out of the refrigerator less than the time frame
Storage Daily
indicated on the cartridge box. If the temperature at which the cartridges are
stored is in doubt, use controls to verify that the cartridges are performing properly.
Check Thermal
i-STAT analyzers contain a thermal control subsystem consisting of two thermal
Control System
probes with thermistors and heating contact wires. When measurements are
Every Six Months
performed at a controlled temperature, the thermal probes in the analyzer contact
the metalized area under the chips in the cartridge and maintain the temperature
of the sensors and the fluids that come into contact with these sensors at the
required temperature ± 0.15°C.
A quality check is performed on the thermal probes each time the external
Electronic Simulator is used. To complete this check, the surface temperature of
the external Electronic Simulator must not fluctuate. If this condition is not met,
the thermal probe check is not completed. Therefore, i-STAT recommends that the
thermal probe check be verified every six months.