3.7:
In the JammLite utility, select the i-STAT 300 Analyzer within the Instrument drop-
down menu.
3.8:
Make sure there is at least one port number listed under the Port Drop-Down List.
Note: If the Port drop-down list says "None." make certain the connection from the Serial
Downloader to the computer is tight. Then reboot the computer and return to step 3.3.
3.9:
Check that the Application and CLEW listings match those in the Product Update for
the current software release. If the update is for CLEW only, select None for Application.
3.10:
Click on the "Update" button.
Note: Application and CLEW numbers are for example only.
3.11:
Follow the onscreen instructions.
Art: 731336-01B
Rev. Date: 13-Aug-13
11