Abbott i-STAT 1 System Manual page 100

Hide thumbs Also See for i-STAT 1:
Table of Contents

Advertisement

PREFERENCE WINDOW:
for
Option
OPERATOR ID
Minimum and maximum allowed operator ID length (scanned
or manually entered)
REPEAT ID ENTRY
Operator must enter ID twice. Handheld prompts operator to
start again if IDs do not match.
INCLUDE ID ON
Enables/Disables printing of operator IDs on printouts from
PRINTOUT
the Martel or i-STAT printer.
BARCODE OPTIONS
The type of barcodes used for Operator ID. See table below.
MANUAL ENTRY CHECK
Options are None, ISBN Modulus 11 Check, and IBM
DIGIT
Modulus 10 Check.
INVALID OPERATOR
Behavior of handheld when Operator ID not in stored list or
certification date expired Options are: Not enabled (continue
without warning), Warn User (prompt to continue), and
Lockout (block testing until a valid Operator ID is scanned/
entered).
EXPIRATION
Allows a System Administrator to define a time period (1-255
NOTIFICATION
days) in which the operator will be notified by a message on
the i-STAT 1 handheld display of their competency expiration
date.
PATIENT ID
Minimum and maximum allowed patient ID length (scanned
or manually entered)
REPEAT ID ENTRY
Operator must enter patient ID twice. Analyzer prompts
operator to start again if IDs do not match.
PATIENT ID RECALL
Operator can recall last patient ID when analyzer prompts for
Patient ID.
BARCODE OPTIONS
The type of barcodes used for Patient ID. See table below.
MANUAL ENTRY CHECK
Options are None, ISBN Modulus 11 Check, and IBM
DIGIT
Modulus 10 Check.
o
p
Id o
perator and
atIent
ptIons
Description
Default
Min = 0 Max = 15
If operator IDs are a fixed length, the min. and max. settings
should both be equal to the ID length.
Enabled: repeat required This option can be set for manual and/or scanned ID Entry.
Enabled
Disabling the printing of operator IDs can prevent uncertified
operators from learning the IDs of certified operators.
All barcode types
None
Check digit algorithms are given in HL7 Specification, Section
2.9.5.3
Continue without
This option should not be enabled if the Use Operator List
warning
option is disabled.
Separate Actions can be chosen for Certification Expired or
Operator Not On List.
Off
See the Technoical Bulletin "Operator Competency Notification
on the i-STAT 1 Handheld" for full details.
Min = 0 Max = 15
If ID numbers are a fixed length, the min. and max. settings
should both be equal to the ID length.
Repeat ID enabled
This option can be set for manual and/or scanned ID entry.
The most recent patient ID is recalled by pressing the → key.
Enabled
All barcode types
None
Check digit algorithms are given in HL7 Specification, Section
2.9.5.3
Comments

Advertisement

Table of Contents
loading