Abbott i-STAT 1 System Manual page 267

Hide thumbs Also See for i-STAT 1:
Table of Contents

Advertisement

2. UPDATING THE i-STAT 1 HANDHELD USING THE CDS VERSION 5
CUSTOMIZATION WORKSPACE WITH A SERIAL OR NETWORK DOWNLOADER
OR DOWNLOADER / RECHARGER, OR DRC-300
2.1:
Before starting the process, make sure all the required equipment / information is
available:
• Computer with CDS Version 5 application
• Access to http://abbottpointofcare.com/softwareupdate
• i-STAT System Equipment
(1) i-STAT 1 handheld
(2) Serial or Network Downloader, Downloader / Recharger, or DRC-300
Note: These instructions assume that the Serial and Network Downloader
types being used for the update process are already installed and in-use at the
customer's facility.
(3) Electronic Simulator
2.2:
Close all open programs on the computer including i-STAT Central Data Station.
2.3:
Navigate to www.abbottpointofcare.com/SoftwareUpdate.
2.4:
Select Language from the drop-down menu.
2.5:
If prompted, indicate the type of Data Management (if any) used with the i-STAT System.
Otherwise, proceed to Step 2.6.
2.6:
When prompted, click on the software file link and save file to:
• Desktop, or
• If you use CDS Version 5, save the file in the c:\bins directory. If the c:\bins directory
does not exist, a bins folder will need to be created under the c: drive.
Close the "Download Complete" window.
2.7:
Navigate to saved file location. Double click the software file "SUXXXXXX.exe".
2.8:
Click Run. If a Command window opens prompting to overwrite, answer "Y" and then
press Enter. Continue answering "Y" to all prompts that appear until the Command window
closes.
2.9:
Update the CLEW and JAMS versions in the CDS Customization Workspace.
• Open the Central Data Station application.
• Click on Main → Open Administration Function → Customization
Art: 731336-01B
Rev. Date: 13-Aug-13
5