Abbott i-STAT 1 System Manual page 249

Hide thumbs Also See for i-STAT 1:
Table of Contents

Advertisement

1. 1 3
By default, the lowest numbered COM port will
automatically be selected.
If the downloader is connected to a different COM
port, change the selection to that COM port now.
JammLite 4.3
Instrument
I-STAT 300 Analyzer
Port
COM1
IP Address
Application
JAMSXXXX.BIN
CLEW
AXX.CLW
1. 1 4
Check that the
match those in the Product Update.
Click the
Update
JammLite 4.3
Instrument
I-STAT 300 Analyzer
Port
COM1
IP Address
Application
JAMSXXXX.BIN
CLEW
AXX.CLW
Note: Application and CLEW numbers are for example only.
The "numbers" have been replaced with X's in the example
above and will change with each software update.
Art: 731335-01B Rev. Date 13-AUG-13
Update
Exit
Application
and
CLEW
button.
Update
Exit
If no ports are displayed, close
all open programs including
JammLite, and then re-launch
JammLite.
If JammLite still has no available
COM ports listed, call your Support
Services representative for assistance.
X
listings
If an error occurs, check the
serial connection between the
downloader and the PC, as well
as the power connection to the
downloader.
If connected correctly, select a
different COM port (Do not select
TCP/IP) within the dropdown
X
menu and click Update.
If errors persist after trying each of
the COM ports listed in JammLite,
verify the serial number of your
downloader and call your Support
Services Representative for
assistance.
7