Procedure (10 Mhz To 500 Mhz) E4401B And E4411B - Agilent Technologies ESA-E Series User Manual

Esa spectrum analyzers, 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Functional Testing
Displayed Average Noise Level

Procedure (10 MHz to 500 MHz) E4401B and E4411B

1. Connect the equipment as shown in
2. Press
,
System
Power On/Preset
analyzer. Wait for the preset routine to finish. Set the analyzer by
pressing the following keys:
,
FREQUENCY
Start Freq
,
Stop Freq
500 MHz
,
AMPLITUDE
–70 dBm
,
AMPLITUDE
More 1 of 2
(75 Ω Input only)
,
BW/Avg
1 MHz
Video BW Auto Man
3. Press the following keys on the analyzer:
Single
,
BW/Avg
Average (On)
Single
Wait until AVG 3 is displayed to the left of the graticule (the
analyzer will take 3 sweeps, then stop).
4. Press
Peak Search
Frequency column (a) in
500 MHz.
5. Press the following keys on the analyzer:
,
Sweep
Sweep (Cont)
,
BW/Avg
Average (Off)
,
BW/Avg
Resolution BW (Auto)
Video BW (Auto)
,
SPAN
50 kHz
FREQUENCY
6. Press
, and set the center frequency of the analyzer to the
Center Freq
frequency recorded in the Measured Frequency column (a) of
Table 3-1
for 10 MHz to 500 MHz.
3-6
Figure
,
Preset (Factory)
,
10 MHz
(50 Ω Input only)
,
,
Amptd Units
dBmV
,
10 kHz
,
,
3
Enter
. Record the measured frequency in the Measured
Table 3-1 on page 3-22
3-1.
,
on the
Preset
,
,
Ref Level
–21.24 dBmV
for 10 MHz to
Chapter 3