ESA Functional Tests
Displayed Average Noise Level
Figure 4-1 Equipment Setup
Figure 4-2 Equipment Setup

Procedure (10 MHz to 500 MHz) E4401B and E4411B E7401A

1. Connect the equipment as shown in
2. Press
,
System
Power On/Preset
preset routine to finish. Set the analyzer by pressing the following keys:
,
FREQUENCY
Start Freq
, 500,
Stop Freq
MHz
,
AMPLITUDE
More
, –70,
AMPLITUDE
,
AMPLITUDE
More
,
AMPLITUDE
Ref Level
,
, 1,
BW/Avg
Res BW
, 10,
Video BW
kHz
3. Press the following keys on the analyzer:
Single
,
BW/Avg
Average
Single
Wait until AVG 3 is displayed to the left of the graticule (the analyzer will take 3
sweeps, then stop).
142
Figure 4-1 Figure
,
(Factory),
Preset Type
, 10,
MHz
,
,
Y Axis Units
dBm
(50 Ω Input only)
dBm
(75 Ω Input only)
,
,
Y Axis Units
dBmV
(75 Ω Input only)
,
–21.24, dBmV
MHz
(On), 3,
Enter
4-2.
on the analyzer. Wait for the
Preset
Chapter 4