h Set the clock mode to Both Edges.

Adjust sampling positions using Eye Finder

1 Select the Sample Positions button. The Eye Finder window will appear.
2 In the "Buses/Signals" section of the Eye Finder window, ensure that the
check box next to "My Bus 1" is checked.
3 Drag the blue bar for "My Bus 1" to approximately -2.8 ns.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 3: Testing Logic Analyzer Performance
Testing the Module Using a 16900-Series Mainframe
43