Adjust Sampling Positions Using Eye Finder - Agilent Technologies 16962A Service Manual

Logic analyzer
Table of Contents

Advertisement

3
Testing Logic Analyzer Performance

Adjust sampling positions using Eye Finder

46
Set the trigger position and acquisition memory depth:
5
Set the Trigger Position to 100% Poststore.
a
Set the Acquisition Depth to 128K.
b
Select the Thresholds and Sample Positions button.
1
The Thresholds and Sample Positions dialog will appear.
2
In the "Buses/Signals" section of the Thresholds and
Sample Positions dialog, ensure that the check box next to
"My Bus 1" is checked.
Drag the blue bar for "My Bus 1" to approximately
3
- 3.6 ns.
16962A Logic Analyzer Service Guide