Section 1: Introduction
180-235 VAC (N. America, Japan, most of Central/South America, etc.):
If 3-phase and single-phase feeds (6-wire):
• Make sure ground wire is connected directly to the frame ground stud on the pedestal floor. Attach no
other wires here.
• Connect L and N @ 200-230 VAC (single-phase)
• Connect 1, 2, and 3 @ 180-235 VAC (3-phase)
If 3-phase with neutral fed from 3-phase (5-wire):
• Make sure ground wire is connected directly to the frame ground stud on the pedestal floor. Attach no
other wires here.
• Connect 1, 2, 3, and N @ 180-235 VAC (3-phase)
• No L available. Connect jumper from 1 to L.
If 3-phase only (4-wire):
• Make sure ground wire is connected directly to the frame ground stud on the pedestal floor. Attach no
other wires here.
• Connect 1, 2, and 3 @ 180-235 VAC (3-phase)
• No L or N available, so the internal outlet will not be functional.
342-456 VAC (Europe, etc.):
 If 3-phase and single-phase feeds (6-wire):
• Make sure ground wire is connected directly to the frame ground stud on the pedestal floor. Attach no
other wires here.
• Connect L and N @ 200-230 VAC (single-phase)
• Connect 1, 2, and 3 @ 342-456 VAC (3-phase)
If 3-phase with neutral fed from 3-phase (5-wire):
• Make sure ground wire is connected directly to the frame ground stud on the pedestal floor. Attach no
other wires here.
• Connect 1, 2, 3, and N @ 342-456 VAC (3-phase)
• No L available. Connect jumper from 1 to L.
 If 3-phase only (4-wire):
This configuration is not possible. Use 5-wire or 6-wire only. Neutral is required. Warning! Ground
(earth) connection is necessary for safety. Never compromise safety by returning current through the
ground.
2-14
CP2000-SB User Manual
020-100162-06 Rev. 1 (03-2014)