3–3 ELECTRICAL SPECIFICATIONS
Electrical conduit, junction boxes, outlets, circuit breakers, and switches should be in place before installing the LO-
GIQ 500 console.
3–3–1 Power Supply
Voltage setup is performed in the factory, since different rear panels which contain different power cables and circuit
breakers are used for the 100 VAC and 220 VAC versions.
3–3–2 Facility Power Receptacle
A separate power outlet with a 20 amp circuit breaker for 115 VAC units, or a 7.5 amp circuit breaker for 220 VAC units,
is recommended. The specific power receptacle used depends on your country's power line standards.
The receptacle should have International Electrotechnical Commission (IEC) approval, or equivalent.
3–4 STORAGE AND OPERATION REQUIREMENTS
The LOGIQ 500 is shipped in a single container excluding PROBES. Shipping weight is approximately 472 lbs
(214kg). The size of the container is B120 cm x W66 cm x H139 cm. (47 in. x 26 in. x 55 in). Table 3–1 provides a
summary of temperature, atmospheric pressure, and humidity tolerances for shipping, installation, and operation:
TEMPERATURE (DEG.C)
ATOMOSPHERIC PRESSURE
HUMIDITY (%)
(Non–condensing)
TABLE 3–1
STORAGE AND OPERATION REQUIREMENTS
PARAMETER
STORAGE
–10 to 60
(DEG.F)
14 to 140
700 to 1060
(hPa)
30 to 95
OPERATION
10 to 40
50 to 104
700 to 1060
30 to 85