6–2–71 HDDB Assy (FRU No. 522)
The HDDB Assy is not used for the console with the 3.5–inch HDD Assy.
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–95.
1.
Turn OFF the system.
2.
Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–41 on page 6–121.
Disconnect three connectors ( 1 – 3 ).
3.
Unscrew four screws ( 4 – 7 ).
4.
5.
Remove the HDDB Assy.
Note
HDDB ASSY DISASSEMBLY
ILLUSTRATION 6–95
6–168
LOGIQ 500 SERVICE MANUAL
P9030TA