6–2–53 Rear CONN Panel Assy (FRU No. 501) and Rear CONN Panel2 Assy (FRU No. 501B)
Time Required
0.5 Hour
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–68 and ILLUSTRATION 6–69.
1.
Turn OFF the system.
2.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
3.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
The Linear Slide Cover is installed from the production unit of LOGIQ500 console with software Ver-
sion 5.00y or later. Refer to ILLUSTRATION 6–30 on page 6–93. Remove the Linear Slide Cover.
Unscrew the two screws ( 1
ILLUSTRATION 6–68. The Rear CONN Spacer is installed from the production unit of LOGIQ500
console with software Version 5.00y or later.
Note
Note
2 ) to remove the Rear CONN Spacer.
and
REAR CONN SPACER DISASSEMBLY
ILLUSTRATION 6–68
6–138
LOGIQ 500 SERVICE MANUAL
P9030TA
Refer to