GE MEDICAL SYSTEMS
REV 14
6–4–4 Installing Software (continued)
5.
Turn the Laptop PC power OFF.
6.
Disconnect the interface cable from the system.
6–4–5 Final Procedures
1.
Turn the system power ON.
2.
Check that the version number of the system software displayed on the CRT monitor screen during the power up
sequence is correct.
3.
Set up the system preset parameters without regard to current settings. For set-up procedures, refer to LOGIQ
500 Proprietary Manual, P9030TB.
3–1 Set the Country/Zone properly. It takes approximately 40 minutes to complete this step.
3–2. Set other three parameters, Video Signal, Power Frequency and Modem Kind.
3–3. Reset the system using System Reset function of the Service Menu. Refer to LOGIQ 500 Proprietary Manu-
al, P9030TB.
4.
Perform functional checks to verify that the system is functioning properly. Refer to Chapter 4, FUNCTIONAL
CHECKS, in this manual.
5.
Load the back-up user data into the system. For the procedure of loading back-up user data, refer to sub-section
4–3–3, Utility Menu in Chapter 4, FUNCTIONAL CHECKS.
6.
This completes the installation of the LOGIQ 500 system software.
6–239
LOGIQ 500 SERVICE MANUAL
P9030TA