6–2–48 Gain Encoder (FRU No. 408)
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–62.
1.
Turn OFF the system.
2.
Remove the Keyboard Panel Assy. Refer to 6–2–41 on page 6–121.
Unscrew four screws ( 1 – 4 ).
3.
Disconnect the gain encoder connector ( 5 ).
4.
The appearance of new Gain Encoder (2206005) is different from the old one (2164871).
The new Gain Encoder (2206005) can be used only for the console with the software version 3.30 or
later.
GAIN ENCODER DISASSEMBLY
ILLUSTRATION 6–62
Note
Note
6–130
LOGIQ 500 SERVICE MANUAL
P9030TA