GE MEDICAL SYSTEMS
REV 14
6–3 FUSE REPLACEMENT
6–3–1 Introduction
LOGIQ 500 has sixteen replaceable fuses. The table below states the items of those fuses.
LOCATION
REAR1 ASSY
TB3
A7SW CONT ASSY
A8 PFC ASSY
INSIDE
POWER
SUPPLY
UNIT
A10 HV REG ASSY
6–3–2 Replacement Procedures
Concerning replacing each fuse of LOGIQ 500, refer to the following steps;
This system contains no operator serviceable parts inside. Please refer servicing to qualified person-
nel from GE. (See Address of Chapter 1.)
Parts Required
FRU
PART NAME
801
Fuse Set
Tools Required
Screwdriver
Hexagonal Wrench (Key Size: 3mm)
TABLE 6–1
REPLACEABLE FUSES
Ref. No.
RATING
F1
1A
F1
800mA
F2
1.6A
F3
5A
F1
500mA
F2
500mA
F1
8A
F1
800mA
F2
350mA
F3
800mA
F4
6.3A
F5
630mA
F6
630mA
F7
350mA
F8
350mA
F9
350mA
Note
QTY
1
6–227
LOGIQ 500 SERVICE MANUAL
REMARKS
250V
250V
250V
500V
For 110V model
250V
250V
250V
250V
250V
250V
125V
250V
250V
250V
250V
250V
PART #
REMARKS
P9522RF
RENEWAL PARTS
P9030TA

Hide quick links: