GE MEDICAL SYSTEMS
REV 14
6–2–21 Side Cover Right (FRU No. 302)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–34.
1.
Turn OFF the system.
2.
Unscrew four screws and remove the Front Base Cover (FRU No. 307). Refer to 6–2–26 on page 6–103.
Remove the five screw caps and unscrew five screws ( 1 – 5 ).
3.
4.
Open the Rear Door and remove the Side Cover Right.
1
2
RIGHT COVER DISASSEMBLY
ILLUSTRATION 6–34
6–98
LOGIQ 500 SERVICE MANUAL
3
302
5
4
307 Front Base Cover
RENEWAL PARTS
P9030TA