GE MEDICAL SYSTEMS
REV 14
6–2–55 Rear 2 Assy (FRU No. 503)
Time Required
1 Hours
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–71 on page 6–143.
1.
Turn OFF the system.
2.
Remove the Rear CONN Panel Assy (FRU 501) or the Rear CONN Panel2 Assy (FRU 501B). Refer to 6–2–53 on
page 6–138.
Unscrew four screws ( 1 – 4 ) to remove the rear plate.
3.
Unscrew ten hexagonal bolts ( 5 – 14 ).
4.
5.
Remove the Rear 2 Assy.
6–142
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS