Breaker 20A (Fru No. 714), Breaker 7.5A (Fru No. 715) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–94 Breaker 20A (FRU No. 714), Breaker 7.5A (FRU No. 715)
Time Required
30 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–124 on page 6–221.
1.
Turn OFF the system.
2.
Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
Unscrew four screws ( 1 – 4 ).
3.
4.
Remove the PS Rear Panel.
Unscrew two screws ( 5 and 6 ).
5.
Unscrew four screws ( 7 – 10 ).
6.
Remove four cables ( 11 – 14 ).
7.
6–220
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS