Probe Environmental Requirements

Operation: Ambient temperature 18° to 30° C
Storage: -10° to 50° C
NOTE:
Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32).
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
NOTICE
-10 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
!! NOTICE:
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
2-2-5

Time and Manpower Requirements

Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Have two people available to deliver and unpack the Voluson® 730Expert.
CAUTION
Attempts to move the unit considerable distances or on an incline by one person could result
!! CAUTION:
in injury or damage or both.
2-2-6

System Specifications

2-2-6-1
Physical Dimensions of Voluson® 730Expert
The physical dimensions of the Voluson® 730Expert unit are summarized in
Table 2-6
Table 2-5
142 cm / 55.9 inches
2-2-6-2

Weight without Monitor and Peripherals

Table 2-6
2-2-6-3

Acoustic Noise Output

max. 57dB(A)
2-2-6-4

Electrical Specifications

Please refer to
3
DRAFT (A
lists the size of Voluson® 730Expert, with monitor and without on-board peripherals.
Physical Dimensions of Voluson® 730Expert
Height
Weight of Voluson® 730Expert with Monitor and without other Peripherals
Model
Voluson® 730Expert
Section 2-2-2-1 "Voluson® 730Expert Power Requirements" on page
Chapter 2 - Site Preparation
29, 2008)
PRIL
Width
68 cm / 26.8 inches
Weight [kg]
136