I/F Board Assy (Fru No. 405) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–45 I/F Board Assy (FRU No. 405)
Time Required
25 Minutes
Tool Required
Screwdriver
Hexagonal Socket Wrench
Procedure
Refer to ILLUSTRATION 6–59.
1.
Turn OFF the system.
2.
Remove the Keyboard Panel Assy. Refer to 6–2–41 on page 6–121.
3.
Remove the VFD Assy. Refer to 6–2–49 on page 6–131.
Unscrew six screws ( 1 – 6 ) and remove the Holding Metal.
4.
Disconnect the three connectors ( 7 – 9 ).
5.
Remove one Stud ( 10 ).
6.
Unscrew seven screws ( 11 – 17 ) and remove the I/F Board Assy.
7.
I/F BOARD ASSY DISASSEMBLY
ILLUSTRATION 6–59
6–125
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents