A5 Ps +5V 30A (Fru No. 705) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–87 A5 PS +5V 30A (FRU No. 705)
Time Required
30 Minutes
Tool Required
Screwdriver
Hexagonal Wrench (Key Size: 3 mm)
Procedure
Refer to ILLUSTRATION 6–117 on page 6–207.
1.
Turn OFF the system.
2.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
3.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
4.
Remove the PS Cover Right. Refer to 6–2–83 on page 6–198.
Slide the transparent cover ( 1 ) upwards and remove it.
5.
Unscrew two screws ( 2 and 3 ).
6.
Remove the four cables ( 4 – 7 ).
7.
Unscrew one hexagonal bolt ( 8 ).
8.
9.
Pull the A2 PS –5.2V 11A outwards.
10. Unscrew two screws ( 9 and 10 ).
11. Remove the four cables ( 11 – 14 ).
12. Disconnect one connector ( 15 ).
6–206
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents