GE MEDICAL SYSTEMS
REV 14
6–2–67 Probe CONN 1 Assy (FRU No. 516)
Time Required
45 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–87.
1.
Turn OFF the system.
2.
Remove the Probe CONN Set Assy (FRU 513). Refer to 6–2–65 on page 6–156.
3.
Remove the Probe CONN 2 Assy (FRU 516). Refer to 6–2–68 on page 6–160.
Unscrew four hexagonal bolts ( 1 – 4 ).
4.
5.
Remove the Probe CONN 1 Assy.
If the customer uses the P509 Probe, the Probe CONN1 Assy for P509 Probe (2180315) is installed.
In this case, you have to disconnect the connector from the RCTL2 Assy before removing Probe
CONN Set Assy.
Î Î
Î Î Î Î Î Î
Î Î Î
Î Î Î Î Î Î Î Î
Î
Î Î Î Î
Note
PROBE CONN 1 ASSY DISASSEMBLY
ILLUSTRATION 6–87
6–159
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS