Dnid2 Assy (Fru No. 620) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–82 DNID2 Assy (FRU No. 620)
The DNID2 Assy is supplied with the optional Single CWD upgrade kit.
Time Required
20 Minutes
Tool Required
Screwdriver
Procedure
Do not bend or flex the boards when mounting/dismounting the piggyback boards. Surface
mount IC boards are very susceptible to damage from flex/torque.
Refer to ILLUSTRATION 6–112 on page 6–195.
1.
Turn OFF the system.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
3.
Unscrew the four Nest Cover screws.
4.
Remove the Nest Cover.
5.
Disconnect the CB16J62 connector from the J2 connector on the DNID2 Assy.
6.
Remove the ANLP2 Assy (FRU 602) installed in slot 4. Refer to 6–2–75, P.C. Board(s) on page 6–180.
Unscrew one screw ( 1 ).
7.
8.
Remove the DNID2 Assy out of the ANLP2 Assy.
Note
CAUTION
6–194
LOGIQ 500 SERVICE MANUAL
P9030TA