GE MEDICAL SYSTEMS
REV 14
6–2–63 ECG Panel Assy (FRU No. 511)
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–80.
The new type of ECG Panel Assy is installed from production unit of LOGIQ500 console with soft-
ware Version 5.00 y or later. Refer to page 6–153.
1.
Turn OFF the system.
2.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
Disconnect three connectors ( 1 – 3 ).
3.
Unscrew two screws ( 4 and 5 ).
4.
Loosen two screws ( 6 and 7 ).
5.
6.
Remove the ECG Panel Assy.
NOTE
ECG PANEL ASSY DISASSEMBLY (a)
ILLUSTRATION 6–80
6–152
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS