6–2–83 A1 PS +5V 32A (FRU No. 701)
Time Required
30 Minutes
Tool Required
Screwdriver
Hexagonal Wrench (Key Size: 3 mm)
Procedure
Refer to ILLUSTRATION 6–113 on page 6–199.
1.
Turn OFF the system.
2.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
3.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
Unscrew five screws ( 1 – 5 ).
4.
5.
Remove the PS Cover Right.
Slide the transparent cover ( 6 ) upwards and remove it.
6.
Unscrew two screws ( 7 and 8 ).
7.
Remove the two cables ( 9 and 10 ).
8.
Unscrew one hexagonal bolt ( 11 ).
9.
10. Pull the A1 PS +5V 32A outwards.
11. Unscrew two screws ( 12 and 13 ).
12. Remove the four cables ( 14 – 17 ).
13. Disconnect one connector ( 18 ).