6–2–64 Nest Fan Assy (FRU No. 512)
Time Required
45 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–83.
1.
Turn OFF the system.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
3.
Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
4.
Remove the Top Cover (FRU 310). Refer to 6–2–29 on page 6–108.
The FRAME BASE is installed from the production unit of the LOGIQ500 console with software ver-
sion 5.00 y or later. Refer to ILLUSTRATION 6–82. Unscrew six screws ( 1 – 6 ) to remove the
FRAME BASE.
1
2
3
4
5
6
NEST FAN ASSY DISASSEMBLY(a)
ILLUSTRATION 6–82
NOTE
6–154
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS