6–2–50 Keyboard Case (FRU No. 410)
Time Required
1 Hour
Tool Required
Screwdriver
Hexagonal Socket Wrench
Procedure
Refer to ILLUSTRATION 6–64.
1.
Turn OFF the system.
2.
Remove the Keyboard Knobs. Refer to 6–2–42 on page 6–122.
3.
Remove the Trackball. Refer to 6–2–43 on page 6–123.
4.
Remove the TGC Assy. Refer to 6–2–44 on page 6–124.
5.
Remove the Gain Encoder. Refer to 6–2–48 on page 6–130.
6.
Remove the VFD Assy. Refer to 6–2–49 on page 6–131.
7.
Remove the I/F Board Assy. Refer to 6–2–45 on page 6–125.
8.
Remove the three LED SW Keyboards with cover metals. Refer to 6–2–47 on page 6–128.
9.
Remove the A/N Keyboard Assy. Refer to 6–2–46 on page 6–126.
KEYBOARD CASE
ILLUSTRATION 6–64
6–132
LOGIQ 500 SERVICE MANUAL
P9030TA