Probe Conn 2 Assy (Fru No. 517) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–68 Probe CONN 2 Assy (FRU No. 517)
Time Required
45 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–88.
1.
Turn OFF the system.
2.
Remove the Probe CONN Set Assy (FRU 513). Refer to 6–2–65 on page 6–156.
Unscrew four screws ( 1 – 4 ).
3.
4.
Remove the Probe CONN 2 Assy.
Î Î Î Î Î Î Î Î Î
Î Î
Î Î Î Î Î
Î Î Î
Î Î
Î
PROBE CONN 2 ASSY DISASSEMBLY
ILLUSTRATION 6–88
6–160
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS