GE MEDICAL SYSTEMS
REV 14
6–2–89 A8 PFC Assy (FRU No. 707)
Time Required
30 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–119 on page 6–211.
1.
Turn OFF the system.
2.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
3.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
4.
Remove the PS Cover Right. Refer to 6–2–83 on page 6–198.
Remove the transparent cover ( 1 ).
5.
Unscrew two screws ( 3 and 4 ).
6.
Remove the two cables ( 5 and 6 ).
7.
Remove the transparent cover ( 2 ).
8.
Unscrew four screws ( 7 – 10 ).
9.
10. Remove the seven cables ( 11 – 17 ).
11. Disconnect one connector ( 18 ).
12. Unscrew two screws ( 19 and 20 ).
6–210
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS