6–2–17 Neck Assy (FRU No. 201)
Time Required
0.6 Hour
Tool Required
Screwdriver
Hex Wrench
Procedure
Refer to ILLUSTRATION 6–29 on page 6–93.
1.
Turn OFF the system and unplug the power cord from outlet.
2.
Lift and set the Monitor to the highest position by pushing down the Up/down Release Button located on the Neck
Assy.
3.
Remove the Monitor Assy from the Neck Assy. Refer to 6–2–1 on page 6–66.
4.
Remove the Neck Space Plate.
5.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
6.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
The Linear Slide Cover is installed from the production unit of LOGIQ500 console with software Ver-
sion 5.00y or later. Refer to ILLUSTRATION 6–30 on page 6–93. Unscrew six screws (
remove the Linear Slide Cover.
7.
Remove the Rear Cover (FRU 303). Refer to 6–2–22 on page 6–99.
8.
Disconnect the three connectors (
The other two connectors are connected with the Mother Board.
Removing the Rear CONN Panel Assy eases to connect and disconnect the cables. For removal of
the Rear CONN Panel Assy, refer to 6–2–53 on page 6–138.
9.
Unscrew two screws and remove two frame ground wires.
10. Remove thirteen hexagonal bolts ( 1 –
Note
14
16
). One of these connectors is connected with the Power Supply Assy.
Note
13
).
6–92
LOGIQ 500 SERVICE MANUAL
P9030TA
17
22