Ventilator Components; Electrical Specifications - GE Aestiva/5 Quick Start Manual

Anesthesia machine
Table of Contents

Advertisement

Ventilator components

Flow transducer
Type:
Variable orifice flow sensor
Dimensions:
22 mm OD and 15 mm ID
Location:
Inspiratory outlet and
expiratory inlet
Optional autoclavable sensor available
Oxygen sensor
Type:
Galvanic fuel cell
Life cycle:
Approximately 18 months
(dependent on usage)
Anesthetic agent delivery
Vaporizers:
Tec 4, Tec 5, Tec 6 Plus, Tec 7
Number of positions:2 or 3
Mounting:
Tool-free installation Selectatec
manifold interlocks and isolates
vaporizers
Tec 7
®
Tec 6 Plus

Electrical specifications

Current leakage
120 V:
< 300µA
Light package
Task light:
12 V, 3 lamps, type 194, .270A each
Goose neck
(optional):
12 V, type 1815, .200A
Power and battery backup
Power input:
120 Vac, 60 Hz, 10A
Backup power:
Demonstrated battery backup time
under typical operating conditions is
45 minutes when fully charged
Battery type:
Internal rechargeable sealed
lead acid
Power cord:
Length: 5 m/16.4 ft
Rating: 15A @ 120 Vac
Communication port
Serial interface:
Isolated RS-232C compatible port
Inlet/outlet modules (
120 V)
System circuit
breakers:
No outlets 5A w/outlets 10A
Outlets (optional):
4 outlets on back, 3-2A,
1-3A individual breakers
and 1-5A combined outlet
breaker, optional isolation
transformer
Auxiliary outlet
box (optional):
5 NEMA outlets on
dovetail-mounted box,
5-2A breakers, isolation
transformer
Tec 6 Plus outlet:
1 IEC 320 located above
vaporizer backbar

Advertisement