Hdd Led Assy (Fru No. 523) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–72 HDD LED Assy (FRU No. 523)
The HDD LED Assy is not available on the console with the new type of 3.5–inch HDD Assy (Quan-
tum FireBall 1280S or newer ).
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–96.
1.
Turn OFF the system.
2.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
3.
Remove the Front Cover (FRU 308). Refer to 6–2–27 on page 6–104.
Unscrew two screws ( 1 – 2 ).
4.
Disconnect one connector ( 3 ).
5.
6.
Remove the HDD LED Assy.
Î Î Î Î Î
Î Î Î Î Î
Î Î Î Î Î
Î Î Î Î Î
Î Î Î Î Î
Î Î Î Î Î
Î
Î Î Î Î Î
Î Î Î Î Î
Î Î Î Î Î
Î Î Î Î Î Î Î Î
Î Î Î
Î Î Î Î Î
Î
Î Î Î Î Î
Î
Note
HDD LED ASSY DISASSEMBLY
ILLUSTRATION 6–96
6–169
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS

Hide quick links: