Section 3-5
Completing the Setup
3-5-1
Purpose of this Section
This section describes how to complete the setup of the Vivid™ S60/Vivid™ S70.
3-5-2
System Specifications
3-5-2-1
System Requirements Verification
•
Verify that the site meets the requirements listed in Chapter 2.
(See:
•
Verify that the specifications below do not conflict with any on-site conditions.
3-5-2-2
Physical Dimensions
Table 3-10 Physical Dimensions of Vivid™ S60/Vivid™ S70 with Monitor and Peripherals
3-5-2-3
Mass with Monitor and Peripherals
Table 3-11
3-5-2-4
Acoustic Noise Level
Less than 55 dB(A) at 20 degrees Celsius, measured in the operators head position, 20 cm in front of
the keyboard's right corner, at 1.30 m above the floor, and in a distance of 1 meter at all four sides, 1
meter above the floor.
3-5-3
Electrical Specifications
WARNING
WARNING
3-5-3-1
Verification of the Vivid™ S60/Vivid™ S70 Voltage Setting
Verify that the mains voltage specified for the Vivid™ S60/Vivid™ S70 is available on-site.
The voltage setting for the Vivid™ S60/Vivid™ S70 is found on a label near the Mains Power Circuit
Breaker on the rear of the Vivid™ S60/Vivid™ S70.
3-5-3-2
Electrical Specifications for the Vivid™ S60/Vivid™ S70
In the table below, the electrical specifications for Vivid™ S60/Vivid™ S70 includes monitor and on
board peripherals.
1
EVISION
Facility Needs on page 2 -
in Transportation Position
Height
Width
104
75
40.95
29.50
Mass of Vivid™ S60/Vivid™ S70 with Monitor, without Probes and Peripherals
Model
Vivid™ S60/Vivid™ S70
Connecting a Vivid™ S60/Vivid™ S70 to the wrong voltage
level will most likely destroy it.
Chapter 3 - System Setup
V
S60N/V
IVID
6.)
Depth
Unit
77
cm
30.30
Inches
Mass [KG]
Mass [LBS]
75
165
S70N
B
S
IVID
ASIC
ERVICE
M
ANUAL