Omem Assy (Fru No. 613) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–79 OMEM Assy (FRU No. 613)
Time Required
20 Minutes
Tool Required
Screwdriver
Procedure
Do not bend or flex the boards when mounting/dismounting the piggyback boards. Surface
mount IC boards are very susceptible to damage from flex/torque.
Refer to ILLUSTRATION 6–109 on page 6–189.
1.
Turn OFF the system.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
3.
Unscrew the four Nest Cover screws.
4.
Remove the Nest Cover.
5.
Remove the TLMC Assy (FRU 606). Refer to 6–2–75 on page 6–180, P.C. Board(s).
Unscrew one screw ( 1 ).
6.
7.
Remove the OMEM Assy out of the TLMC Assy.
CAUTION
6–188
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS