6–2–29 Top Cover (FRU No. 310)
Time Required
30 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–43 on page 6–109.
1.
Turn OFF the system.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
3.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
4.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
5.
Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
Unscrew six screws ( 1 – 6 ).
6.
Unscrew two Hexagonal bolts ( 7 – 8 ).
7.
8.
Remove the Top Cover.
6–108
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS