6–5 POWER SUPPLY REPLACEMENT
6–5–1 Introduction
This section describes the procedures for replacing the power supply unit of LOGIQ 500.
6–5–2 Replacement Procedure
READ and UNDERSTAND these instructions thoroughly before proceeding with the replacement. Perform each step
in sequence and check it off when completed. If a problem occurs after completion, go back through the procedure and
check for implementation errors before contacting your Ultrasound Region Engineer.
Possible operational damage. Failure to strictly follow ESD (Electrostatic Discharge) precau-
tions during this upgrade may cause constant or intermittent operational abnormalities.
Strictly follow all precautions.
Check each step as it is completed.
Parts Required
FRU
PART NAME
717
Power Supply AC110V
718
Power Supply AC220V
Time Required
Approximately four (4) hours
Tools Required
Standard field Service Tool kit
Three Jacks
Pallet (Height: 70mm, Width and Depth: suitable length)
CAUTION
QTY
PART #
1
P9523DB
1
P9523DA
6–243
LOGIQ 500 SERVICE MANUAL
P9030TA
REMARKS
for 110–120V model
for 220–240V model
RENEWAL PARTS