6–2–90 A10 HV REG Assy (FRU No. 708)
Time Required
30 Minutes
Tool Required
Screwdriver
Hexagonal Wrench (Key Size: 3 mm)
Procedure
Refer to ILLUSTRATION 6–120 on page 6–213.
1.
Turn OFF the system.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
Unscrew five screws ( 1 – 5 ).
3.
4.
Remove the PS Cover Left.
Unscrew one screw ( 6 ).
5.
Pull the HV REG Assy outwards and disconnect seven connectors ( 7 – 13 ).
6.
6–212