6–2–47 LED SW Keyboard Set (FRU No. 407)
Time Required
40 Minutes
Tool Required
Screwdriver
Hexagonal Socket Wrench
Procedure
Refer to ILLUSTRATION 6–61 on page 6–129.
1.
Turn OFF the system.
2.
Remove the Keyboard Panel Assy. Refer to 6–2–41 on page 6–121.
3.
Remove the I/F Board Assy. Refer to 6–2–45 on page 6–125.
Remove three Studs ( 1 – 3 ).
4.
Unscrew one screw ( 4 ) and remove the LED SW KEYBOARD (A) with the cover metal from the Keyboard Case.
5.
Unscrew two screws ( 5 and 6 ) and remove the LED SW KEYBOARD (A) from the cover metal.
6.
Disconnect one connector ( 7 ).
7.
Unscrew four screws ( 8 – 11 ) and remove the LED SW KEYBOARD (B) with the cover metal from the Key-
8.
board Case.
Unscrew four screws ( 12 – 15 ) and remove the LED SW KEYBOARD (B) from the cover metal.
9.
10. Unscrew four screws ( 16 – 19 ) and remove the Holding Metal.
11. Unscrew thirteen (13) screws ( 20 – 32 ) and remove the LED SW KEYBOARD (C) with the cover metal from the
Keyboard Case.
12. Unscrew eleven (11) screws ( 33 – 43 ) and remove the LED SW KEYBOARD (C) from the cover metal.
6–128