Ps Fan Set (Fru No. 713) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–93 PS Fan Set (FRU No. 713)
Time Required
30 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–123 on page 6–219.
1.
Turn OFF the system.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
3.
Remove the PS Cover Left. Refer to 6–2–83 on page 6–198.
Unscrew two screws ( 1 and 2 ).
4.
Disconnect one connector ( 3 ).
5.
6.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
7.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
8.
Remove the PS Cover Right. Refer to 6–2–90 on page 6–212.
Unscrew two screws ( 4 and 5 ).
9.
10. Disconnect one connector ( 6 ).
6–218
LOGIQ 500 SERVICE MANUAL