6–2–73 Power Switch Assy (FRU No. 525)
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–97.
The new type of Power Switch Assy is installed from production unit of LOGIQ500 console with soft-
ware version 5.00 y or later. Refer to ILLUSTRATION 6–98 on page 6–171.
1.
Turn OFF the system.
2.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
3.
Remove the Front Cover (FRU 308). Refer to 6–2–27 on page 6–104.
Disconnect one connector ( 1 ).
4.
Unscrew two screws ( 2 and 3 ).
5.
6.
Remove the Power Switch Assy.