6–2–70 HDD Assy (FRU No. 520), HDD Holder Assy (FRU No. 521)
Time Required
1.5 Hours
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–90 on page 6–163.
1.
Make sure the customer data is saved on disks (floppy disks for system software version 1.20 or former console
and MO disk for 2.00 or later console). Refer to 4–3–3 (D) on page 4–21, USER DATA BACK UP.
2.
Turn OFF the system.
3.
Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–41 on page 6–121.
4.
Remove the HDDB Assy (FRU 522). Refer to 6–2–71 on page 6–168.
Unscrew two screws ( 1 and 2 ).
5.
6.
Remove the HDD Holder Assy.
Unscrew four screws ( 3 – 6 ).
7.
8.
Remove the HDD (FRU 520) out of the HDD Holder Assy (FRU 521).
After replacing a HDD Assy with new one, the following works have to be done:
a. Set up the system preset parameters without regard to current settings. For set-up procedures, refer to
LOGIQ 500 Proprietary Manual, P9030TB.
a–1. Set the Country/Zone properly. It takes approximately 40 minutes to complete this step.
a–2. Set other three parameters, Video Signal, Power Frequency and Modem Kind.
a–3. Reset the system using System Reset function of the Service Menu. Refer to LOGIQ 500 Proprietary
Manual, P9030TB.
b. Install the customer data into new HDD using the back-up disk(s) after resetting the system. Refer to 4–3–3
(D) on page 4–21, USER DATA BACK UP.