6–2–61 HV Fan Set (FRU No. 509)
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–77.
1.
Turn OFF the system.
2.
Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
3.
Remove the HV Assy (FRU 508). Refer to 6–2–60 on page 6–148.
Unscrew six screws ( 1 – 6 ) to disassemble the HV Assy.
4.
Disconnect two connectors ( 7 and 8 ).
5.
Unscrew four screws ( 9 – 12 ).
6.
HV FAN SET DISASSEMBLY
ILLUSTRATION 6–77
6–149
LOGIQ 500 SERVICE MANUAL
P9030TA