1.6.3 Connector Symbols

1.7 Electrical, Mechanical & Environmental Specifications
NIBP
CUFF PRESSURE RANGE:
DEFAULT TARGET
CUFF INFLATION:
TARGET CUFF INFLATION
ADJUSTMENT RANGE:
BLOOD PRESSURE
MEASUREMENT RANGE:
Adult / Pediatric:
Neonate:
BLOOD PRESSURE ACCURACY:
BLOOD PRESSURE
DETERMINATION TIME:
PULSE RATE DETERMINATION:
PULSE RATE ACCURACY:
OVERPRESSURE CUT-OFF:
Predictive Temperature Determination (TS & T models)
RANGE:
PROBE ACCURACY:
External Alarm Connector
External Comms Port Connector
Fuse
External Power
Adult / Pediatric:
Neonate:
Adult / Pediatric:
Neonate:
Adult / Pediatric:
Neonate:
Systolic
mmHg
30 - 245
40 - 140
Meets AAMI/ANSI SP-10 Electronic or Automated
Sphygmomanometer Standard for Accuracy. (AAMI/ANSI
standard :± 5 mmHg mean error; 8 mmHg standard deviation;
intra-arterial method.)
20 seconds to 45 seconds typical
Adult:
Neonate:
Adult / Pediatric:
Neonate:
± 3.5 percent
Adult / Pediatric
Neonate
42.2 ? Celsius
Max.
Min
31.1? Celsius
? 0.1
o
C
0 mmHg to 290 mmHg
0 mmHg to 140 mmHg
178 ± 15 mmHg
110 ± 15 mmHg
100 to 250 mmHg; 5 mmHg increments
100 to 140 mmHg; 5 mmHg increments
MAP mmHg
15 - 215
30 - 115
120 seconds maximum,
85 seconds maximum.
30 - 200 BPM
30 - 200 BPM
300 - 330 mmHg
150 - 165 mmHg
108.0° Fahrenheit
88.0° Fahrenheit
o
± 0.2
F
Diastolic
mmHg
10 - 195
20 - 100
11