GE MEDICAL SYSTEMS
REV 5
4–2–1 Basic Controls (Continued)
Step
Check
11.
Press M key.
12.
Roll trackball.
13.
Press Scan Area button.
14.
Press M/D Cursor button.
15.
Click paddle under the 2nd
softkey choice to adjust M
Edge Enhancement up/down.
You can select several types of display format by using the Preset Menu. For the Preset Menu, refer to
Customizing Your System in the LOGIQ 500 User Manual.
Expected result
The M mode timeline should appear below the B image
as shown in ILLUSTRATION 4–2. Whether it takes half
the screen or two–thirds depends on the preset.
The Mode cursor should follow trackball movement and
the timeline should update for new location of focus.
The M mode cursor should be inactive. The scan area
frame should appear. The scan area frame grows wider
or narrower depending on trackball movement.
The M mode cursor should be active again.
Changes the M image.
M–MODE DISPLAY SCREEN
ILLUSTRATION 4–2
Note
4–5
LOGIQ 500 SERVICE MANUAL
P9030TA
FUNCTIONAL CHECKS