6–2–74 MODD (FRU No. 540), MODD Fan (FRU No. 541), MODD Holder Assy (FRU No. 542)
Time Required
30 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–99 on page 6–173.
1.
Turn OFF the system.
2.
Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–41 on page 6–121.
3.
Disconnect the MODD Fan Power Cable.
Unscrew two screws ( 1 and 2 ).
4.
5.
Remove the MODD Fan Assy
6.
Remove the HDDB Assy (FRU 522). Refer to 6–2–71 on page 6–168.
The HDDB Assy is not used for the console with the 3.5–inch HDD Assy.
Unscrew four screws ( 3 – 6 ).
7.
8.
Remove the upper metal plate.
9.
Disconnect the SCSI Cable and MODD Power Cable.
10. Remove the MODD Holder Assy.
11. Unscrew four screws ( 7 –
12. Remove the MODD.
Note
10
).
6–172
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS