GE MEDICAL SYSTEMS
REV 14
6–2–84 A2 PS –5.2V 11A (FRU No. 702)
Time Required
30 Minutes
Tool Required
Screwdriver
Hexagonal Wrench (Key Size: 3 mm)
Procedure
Refer to ILLUSTRATION 6–114 on page 6–201.
1.
Turn OFF the system.
2.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
3.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
4.
Remove the PS Cover Right. Refer to 6–2–83 on page 6–198.
Slide the transparent cover ( 1 ) upwards and remove it.
5.
Unscrew six screws ( 2 – 7 ).
6.
Remove the eight cables ( 8 – 15 ).
7.
Unscrew one hexagonal bolt ( 16 ).
8.
6–200
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS