6–2–66 Shield Finger Long (FRU No. 514), Shield Finger Short (FRU No. 515)
Time Required
45 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–86.
1.
Turn OFF the system.
2.
Remove the Probe CONN Set Assy (FRU 513). Refer to 6–2–65 on page 6–156.
Unscrew eight screws ( 1 – 8 ).
3.
4.
Remove the connector frame.
5.
Remove the Shield Finger Long (FRU 514) or Shield Finger Short (FRU 515).
Î Î Î Î Î Î Î Î Î Î
Î Î
Î Î Î
Î Î Î
Î Î
Î
SHIELD FINGER DISASSEMBLY
ILLUSTRATION 6–86
6–158
LOGIQ 500 SERVICE MANUAL
P9030TA