Nest Mother Assy (Fru No. 614) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–80 Nest Mother Assy (FRU No. 614)
Time Required
2 Hours
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–110 on page 6–192.
An electrostatic discharge may damage a component. Turn OFF power and wear the wrist
strap before you remove a circuit boards. Do not unplug the power cord to keep ground conti-
nuity.
Do not bend or flex the boards when mounting/dismounting each board. Surface mount IC
boards are very susceptible to damage from flex/torque.
1.
Turn OFF the system.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
3.
Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
4.
Remove the Side Cover Right (FRU 302). Refer to 6–2–21 on page 6–98.
5.
Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
6.
Remove the Front Cover (FRU 308). Refer to 6–2–27 on page 6–104.
7.
Remove the Top Cover (FRU 310). Refer to 6–2–29 on page 6–108.
8.
Unscrew the four Nest Cover screws.
9.
Remove the Nest Cover.
10. Remove all the P.C. Boards. Refer to 6–2–75 on page 6–180.
CAUTION
CAUTION
6–190
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS