Inlet Unit 120V, Inlet Unit 220V (Fru No. 720) - GE LOGIQ 500 Service Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 14
6–2–95 Inlet Unit 120V, Inlet Unit 220V (FRU No. 720)
Time Required
30 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–125 on page 6–223.
1.
Turn OFF the system.
2.
Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
3.
Remove the Power Supply Unit. Refer to 6–5–2 on page 6–243.
Unscrew four screws ( 1 – 4 ). Refer to ILLUSTRATION 6–124 on page 6–221.
4.
5.
Remove the transparent plastic cover.
Unscrew two screws ( 5 and 6 ) and remove the two cables.
6.
Unscrew one screw ( 7 ).
7.
8.
Pull the Ground Wire out by hand.
Unscrew two screws ( 8 and 9 ) and remove the two cables.
9.
10. Cut two Tie Wraps and remove them.
11. Remove the Inlet Unit.